Price
$577.00 CAD
Stock
25+ In Stock

  • Pressure Type - ± Plus or Minus
  • Output - 4-20 mA
  • Pressure Range - Inches WC: 0 to ± 1
  • Temperature Range - -1° to 65° C (30° to 150° F)
  • Accuracy - ± 0.073 % Full Scale Accuracy

Specifications

Accuracy

  • < ± 0.14 % of Full Scale

Applications

  • Clean Rooms
  • Energy/Water Management
  • Environmental Testing
  • Leak Detection
  • Medical Instruments

Approvals & Certifications

  • CE

Compensated Temperature

  • -1° to 65° C (30° to 150° F)

Consumption

  • < 30 mA

Electrical Connectors

  • 2 Foot Cable Standard

Loop Resistance

  • Maximum
  • Minimum (Vdc) = 17 + 0.02 x Resistance of Receiver Plus Line

Media

  • Compatible Non-Conducting Air
  • Compatible Non-Conducting Gas

NACE & NIST

  • NIST

NPT Connection

  • 1/8”-27 Female NPT Internal

Operating Temperature

  • -18° to 71° C (0° to 160° F)

Output

  • 0-5 Vdc for Voltage Models
  • 4-20 mA for Current Models

Power Requirement

  • 17-42 Vdc for Current Models
  • 22-30 Vdc for Voltage Models

Pressure Range

  • bar: 0 to 0.34
  • Inches WC: 0 to 15
  • kPa: 0 to 34.47
  • psid: 0 to 5

Pressure Type

  • Differential

Response Time

  • < 5 milliseconds

Sensor Orientation

  • Pressure Port 90° Parallel to Ground

Span & Zero Effect

  • Zero & Span Output: Factory Set to within ± 0.07 mA

Stability

  • < ± 0.1 % of Full Scale for 6 Months at 21° C (70° F)

Storage Temperature

  • -53° to 121° C (-65° to 250° F)

Temperature Effect

  • < ± 1.0 % of Full Scale/38° C (100° F)
show more